ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19784
- Event Type
- Malfunction
- Date Received
- July 11, 2014
- Report Date
- July 4, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
FOLLOW-UP #1: DATE OF SUBMISSION 3/20/15 DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 03/10/2015 WITH THE FOLLOWING FINDINGS: BLACK BOX INFORMATION FROM DATE OF EVENT HAS BEEN OVERWRITTEN. UNABLE TO CONFIRM OR DUPLICATE THE COMPLAINT DURING INVESTIGATION. PUMP POWERS WITH RETURNED BATTERY CAP. BATTERY CAP IS ABLE TO FULLY TIGHTEN. BATTERY COMPARTMENT INTACT. NO ALARMS OBSERVED IN ALARM HISTORY. CURRENT DRAWS WITHIN SPECIFICATIONS. NO "BATTERY LIFE" ISSUES DUPLICATED DURING THE INVESTIGATION. REMOVED PUMP CASE. NO EVIDENCE OF MOISTURE OR LOOSE COMPONENTS INSIDE PUMP. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A POWER (BATTERY LIFE) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 407550 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |