HOMECHOICE
Report
- Report Number
- 1416980-2014-22380
- Event Type
- Malfunction
- Date Received
- July 11, 2014
- Date of Event
- May 9, 2014
- Report Date
- June 18, 2014
- Manufacturer
- BAXTER HEALTHCARE - LARGO
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE DEVICE WAS RECEIVED BY BAXTER FOR EVALUATION. REVIEW OF THE EVENT LOG FOUND EVIDENCE OF THE REPORTED IIPV CONDITION. ALL RETURNED HOMECHOICE DEVICES RECEIVE A RETURNED INSTRUMENT TESTING EVALUATION (RITE). THIS EVALUATION INCLUDES FUNCTIONAL AND ELECTRICAL TESTING ON THE DEVICE. THE CAUSE WAS DETERMINED TO BE A USE ERROR, TIDAL TOTAL UF REMOVAL SET TOO LOW. HOMECHOICE APD SYSTEMS TRAINER¿S GUIDE GIVES INSTRUCTIONS ON HOW TO SET THE TIDAL THERAPY SETTINGS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
DURING EVALUATION OF A RETURNED HOMECHOICE DEVICE, ONE INCREASED INTRA-PERITONEAL VOLUME (IIPV) EVENT WAS IDENTIFIED WHICH OCCURRED IN THE THERAPY INITIATED ON (B)(6) 2014 21:50:07. DURING NIGHT DRAIN CYCLE FIVE, THE PATIENT'S ULTRAFILTRATION READING WAS 1971ML, INDICATING THE HOME PATIENT (HP) DRAINED 1821ML MORE THAN THEIR MAXIMUM PROGRAMMED FILL VOLUME OF 1500ML. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 406219 | HOMECHOICE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - LARGO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR |