MDR report key: 3930005 · Received July 11, 2014

Report

Report Number
2124215-2014-09739
Event Type
Injury
Date Received
July 11, 2014
Date of Event
April 2, 2014
Report Date
May 22, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWP
PMA / PMN Number
P010012
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

(B)(4). UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. IMPEDANCE TESTING WAS COMPLETED AND ALL MEASUREMENTS WERE WITHIN NORMAL LIMITS. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH OUT OF RANGE PACING LEAD IMPEDANCE MEASUREMENTS GREATER THAN 2,000 OHMS. SHOCK PACE IMPEDANCE WAS STEADY AND THRESHOLD VALUE WAS 4.5 VOLTS AT 1.5 MS. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) DISCUSSED THAT THIS COULD BE A SIGN OF A LEAD ISSUE AND RECOMMENDED BRINGING IN THE PATIENT TO SEE IF NOISE CAN BE REPRODUCED. ADDITIONAL INFORMATION WAS RECEIVED INDICATING THAT A REVISION PROCEDURE WAS PERFORMED. THE ICD WAS EXPLANTED AND REPLACED WITH A SUBCUTANEOUS IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (S-ICD) WHILE THE RV LEAD WAS SURGICALLY ABANDONED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
407733 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWP GUIDANT CRM CLONMEL IRELAND E102

Patients

Seq Age Sex Outcome Treatment
1 34 YR Hospitalization| L| R 0175| E102