FDA Adverse Event
Injury
Summary report: N
SUMMIT COLLIMATOR
MDR report key: 3929691
·
Received June 17, 2014
Report
- Report Number
- 3929691
- Event Type
- Injury
- Date Received
- June 17, 2014
- Date of Event
- June 5, 2014
- Report Date
- June 9, 2014
- Manufacturer
- SUMMIT INDUSTRIES
- Product Code
- KPW
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- User Facility report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Description of Event or Problem · 1
FACILITY INFORMED US (SERVICE PROVIDER) THAT THE COLLIMATOR CAME LOOSE AND FELL STRIKING A PT IN THE LEG WHILE ON THE X-RAY TABLE. WE REPAIRED COLLIMATOR FOR FURTHER USE. CONDITION OF PT IS UNK. WE (SERVICE PROVIDER), WERE CALLED IN TO COME FIX THE ISSUE WHEN WE WERE TOLD ABOUT THE COLLIMATOR STRIKING A PT. WE WERE NOT THE CURRENT SERVICE PROVIDER OF THE MEDICAL X-RAY EQUIPMENT. OUR SERVICE ENGINEER REMEMBERED THE RECALL FROM 2007 AND REPAIRED ACCORDING TO THE RECALL INSTRUCTIONS. ORIGINAL RECALL DATED 8/3/2007. THIS REPORT PROVIDED BY: (B)(6). (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 353757 | SUMMIT COLLIMATOR | D800 COLLIMATOR | KPW | SUMMIT INDUSTRIES | D800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |