FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 3926538
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15716
- Event Type
- Injury
- Date Received
- July 10, 2014
- Date of Event
- May 7, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KY, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
.
Description of Event or Problem · 1
IT WAS REPORTED THAT UPON A CHEST X-RAY, IT WAS NOTED THAT THE ATRIAL LEAD HAD DISLODGED. THE LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403066 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1888TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention |