FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 3926408
·
Received July 10, 2014
Report
- Report Number
- 2531779-2014-19532
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Report Date
- June 30, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/30/2014 WITH THE FOLLOWING FINDINGS: KEYPAD COVER IS PEELING OFF THE BASE. ALL KEYPAD BUTTONS RESPOND NORMALLY. THE DISPLAY IS DIM/FADING/REDDISH. THE KEYPAD COVER WAS REMOVED AND NO CONTAMINATION WAS FOUND UNDER CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. EVALUATION REVEALED DISPLAY IS DIM/FADING/REDDISH. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/30/2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403989 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 47 YR |