EON MINI
Report
- Report Number
- 1627487-2014-01373
- Event Type
- Injury
- Date Received
- June 2, 2014
- Date of Event
- May 9, 2014
- Report Date
- May 9, 2014
- Manufacturer
- ST. JUDE MEDICAL ¿ NEUROMODULATION
- Product Code
- GZB
- PMA / PMN Number
- P010032
- Removal / Correction Number
- 1627487-07262012-002-R
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
CORRECTION NUMBER: 1627487-07262012-001-C. THIS IPG SERIAL NUMBER WAS INCLUDED IN FIELD ADVISORIES AND A FIELD CORRECTION. SJM HAS LIMITED INFORMATION RELATED TO THE PATIENT'S MEDICAL HISTORY AND IS UNABLE TO FORM AN OPINION AS TO THE RELEVANCY OF THE PATIENT'S HISTORY TO THE EVENT REPORTED. SJM DEFERS TO THE PATIENT'S PHYSICIAN REGARDING MEDICAL HISTORY.
DEVICE 1 OF 2 REFERENCE MFR REPORT# 1627487-2014-01374. IT WAS REPORTED THE PATIENT EXPERIENCED UNCOMFORTABLE HEATING WHILE RECHARGING. SURGICAL INTERVENTION WAS UNDERTAKEN ON (B)(6) 2014 AND HT IPG WAS EXPLANTED. THE ISSUE RESOLVED. ON AUGUST 1, 2012 ST. JUDE MEDICAL, NEUROMODULATION DIVISION, SENT FIELD ACTION LETTERS TO PATIENTS RELATED TO HEATING WHILE CHARGING AND RAISED AWARENESS OF THIS ISSUE TO PATIENTS. AN INCREASE IN PRIOR NON-REPORTED HEATING WHILE CHARGING EVENTS AND OTHER NON-REPORTED EVENTS WAS EXPECTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 322218 | EON MINI | SCS IPG | GZB | ST. JUDE MEDICAL ¿ NEUROMODULATION | 3788 | 3249788 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Other | IMPLANT DATE:| SCS LEAD: MODEL 3228 |