FDA Adverse Event Injury Summary report: N

TENDRIL SDX

MDR report key: 3926218 · Received July 10, 2014

Report

Report Number
2017865-2014-15539
Event Type
Injury
Date Received
July 10, 2014
Date of Event
April 22, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

FINAL ANALYSIS FOUND INSULATION ABRASION EXPOSING THE OUTER COIL AT 49.4 CM TO 50.3 CM AND 50.8 CM TO 51.3 CM FROM THE CONNECTOR PIN. THE ABRASION IS CONSISTENT WITH THAT OF EXPOSURE TO FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED NOISE. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404671 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1688T/52 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention