TENDRIL SDX
MDR report key: 3926218
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15539
- Event Type
- Injury
- Date Received
- July 10, 2014
- Date of Event
- April 22, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
FINAL ANALYSIS FOUND INSULATION ABRASION EXPOSING THE OUTER COIL AT 49.4 CM TO 50.3 CM AND 50.8 CM TO 51.3 CM FROM THE CONNECTOR PIN. THE ABRASION IS CONSISTENT WITH THAT OF EXPOSURE TO FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE ATRIAL LEAD EXHIBITED NOISE. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 404671 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1688T/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |