FDA Adverse Event
Malfunction
Summary report: N
TENDRIL STS
MDR report key: 3926054
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15395
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- April 18, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. . NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION AT 9.8 CM TO 10.0 CM FROM THE CONNECTOR PIN. THE ABRASION IS CONSISTENT WITH THAT OF EXPOSURE TO FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403789 | TENDRIL STS | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 2088TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |