FDA Adverse Event
Injury
Summary report: N
TENDRIL STS
MDR report key: 3926030
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15418
- Event Type
- Injury
- Date Received
- July 10, 2014
- Date of Event
- April 2, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. FINAL ANALYSIS FOUND CLAVICLE CRUSH DAMAGING OUTER COIL AND INSULATION ABRASION AT 17.0 CM TO 17.8 CM FROM THE CONNECTOR PIN; CAUSING AN ELECTRICAL SHORT.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOW IMPEDANCE, OVERSENSING, LOSS OF CAPTURE. RIB CLAVICLE CRUSH DAMAGE WAS NOTED. THE LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2014.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 402749 | TENDRIL STS | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 2088TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 77 YR | Required Intervention | PM2210, 7455448 |