FDA Adverse Event Injury Summary report: N

TENDRIL STS

MDR report key: 3926030 · Received July 10, 2014

Report

Report Number
2017865-2014-15418
Event Type
Injury
Date Received
July 10, 2014
Date of Event
April 2, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. FINAL ANALYSIS FOUND CLAVICLE CRUSH DAMAGING OUTER COIL AND INSULATION ABRASION AT 17.0 CM TO 17.8 CM FROM THE CONNECTOR PIN; CAUSING AN ELECTRICAL SHORT.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD EXHIBITED LOW IMPEDANCE, OVERSENSING, LOSS OF CAPTURE. RIB CLAVICLE CRUSH DAMAGE WAS NOTED. THE LEAD WAS EXPLANTED AND REPLACED ON (B)(6) 2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402749 TENDRIL STS PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 2088TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 77 YR Required Intervention PM2210, 7455448