FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 3925967
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15326
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- May 21, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT ONLY THE DISTAL PORTION OF THE LEAD WAS RECEIVED. INSULATION ABRASION EXPOSING THE OUTER COIL WAS NOTED AT 6.5 CM TO 6.9 CM FROM THE DISTAL TIP. THE ABRASION IS CONSISTENT WITH THAT OF EXPOSURE TO FRICTION WITH ANOTHER IMPLANTABLE DEVICE. RELIABILITY LABORATORY TECHNICIAN. RELIABILITY LABORATORY TECHNICIAN. ST. JUDE MEDICAL CRMD RELIABILITY LABORATORY.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.