FDA Adverse Event
Malfunction
Summary report: N
TENDRIL SDX
MDR report key: 3925819
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15661
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- April 2, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT WAS RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASIONS AT 46.5 CM - 46.6 CM, 46.8 CM TO 47.1 CM AND 51.3 CM - 51.8 CM FROM THE CONNECTOR PIN. THE ABRASIONS WERE CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE. RELIABILITY LABORATORY TECHNICIAN RELIABILITY LABORATORY TECHNICIAN ST. JUDE MEDICAL CRMD RELIABILITY LABORATORY
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.