FDA Adverse Event Injury Summary report: N

QUARTET

MDR report key: 3925794 · Received July 10, 2014

Report

Report Number
2017865-2014-15262
Event Type
Injury
Date Received
July 10, 2014
Date of Event
May 7, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
OJX
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT PRESENTED IN CLINIC FOR ROUTINE FOLLOW-UP. UPON INTERROGATION THE LEFT VENTRICULAR LEAD EXHIBITED LOSS OF CAPTURE. DURING PRE-OPERATION CHECK, FLUOROSCOPY CONFIRMED THAT THE LEAD HAD DISLODGED INTO THE ATRIUM. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404988 QUARTET PERMANENT PACEMAKER ELECTRODE OJX ST. JUDE MEDICAL, INC., CRMD 1458Q/86 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR Required Intervention