FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 3925759
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15239
- Event Type
- Injury
- Date Received
- July 10, 2014
- Date of Event
- May 15, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.
Description of Event or Problem · 1
IT WAS REPORTED THAT A NON-DEPENDENT PATIENT PRESENTED IN THE HOSPITAL DUE TO MUSCLE STIMULATION. THE VENTRICULAR LEAD EXHIBITED LOW IMPEDANCE IN BIPOLAR CONFIGURATION. NOISE AND HIGH PACING THRESHOLD WERE ALSO NOTED. AN INSULATION ANOMALY WAS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 404920 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1888TC/46 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 86 YR | Required Intervention |