FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 3925759 · Received July 10, 2014

Report

Report Number
2017865-2014-15239
Event Type
Injury
Date Received
July 10, 2014
Date of Event
May 15, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A NON-DEPENDENT PATIENT PRESENTED IN THE HOSPITAL DUE TO MUSCLE STIMULATION. THE VENTRICULAR LEAD EXHIBITED LOW IMPEDANCE IN BIPOLAR CONFIGURATION. NOISE AND HIGH PACING THRESHOLD WERE ALSO NOTED. AN INSULATION ANOMALY WAS SUSPECTED. THE LEAD WAS CAPPED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404920 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1888TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1 86 YR Required Intervention