FDA Adverse Event
Malfunction
Summary report: N
TENDRIL ST
MDR report key: 3925716
·
Received July 10, 2014
Report
- Report Number
- 2017865-2014-15618
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- April 9, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT A PARTIAL LEAD WAS RETURNED. THE LEAD INSULATION WAS ABRADED AT 12.3CM TO 13.1CM FROM THE CUT DISTAL END. THE ABRASION WAS CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE. (B)(6).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 404434 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1788TC/46 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |