FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 3925716 · Received July 10, 2014

Report

Report Number
2017865-2014-15618
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
April 9, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT A PARTIAL LEAD WAS RETURNED. THE LEAD INSULATION WAS ABRADED AT 12.3CM TO 13.1CM FROM THE CUT DISTAL END. THE ABRASION WAS CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE. (B)(6).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404434 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1788TC/46 NA

Patients

Seq Age Sex Outcome Treatment
1