FDA Adverse Event
Injury
Summary report: N
840 VENTILATOR
MDR report key: 3925688
·
Received June 17, 2014
Report
- Report Number
- 8020893-2014-01446
- Event Type
- Injury
- Date Received
- June 17, 2014
- Date of Event
- May 16, 2014
- Report Date
- May 21, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE UNIT, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB), UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MANUFACTURING SPECIFICATIONS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN THE USA THAT DURING PATIENT USE, AN 840 VENTILATOR STOPPED CYCLING. THERE WAS NO REPORTED PATIENT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 356472 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR PURITAN BENNETT IRELAND | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |