FDA Adverse Event Malfunction Summary report: N

CORE IMPACTION DRILL

MDR report key: 3925647 · Received July 10, 2014

Report

Report Number
0001811755-2014-02453
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
June 13, 2014
Report Date
June 13, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
DZJ
PMA / PMN Number
K040369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FAILURE ANALYSIS IS IN PROGRESS; ADDITIONAL INFORMATION WILL BE SUBMITTED ONCE THE QUALITY INVESTIGATION IS COMPLETED.

Additional Manufacturer Narrative · 1

THE REPORTED EVENT WAS DUPLICATED. UPON DISASSEMBLY THE DEVICE WAS CONFIRMED TO HAVE BRINELLED DRIVE SHAFT BEARINGS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES. FOLLOWING THE PROCEDURE IT WAS NOTED THAT THERE WERE BLACK BURN MARKS ON THE BUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A PROCEDURE AT THE USER FACILITY, THE CORE IMPACTION DRILL WAS OVERHEATING. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY, NO MEDICAL INTERVENTION, AND NO ADVERSE CONSEQUENCES. FOLLOWING THE PROCEDURE IT WAS NOTED THAT THERE WERE BLACK BURN MARKS ON THE BUR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404242 CORE IMPACTION DRILL DRIVER, WIRE, AND BONE DRILL, MANUAL DZJ STRYKER INSTRUMENTS-KALAMAZOO

Patients

Seq Age Sex Outcome Treatment
1 UNKNOWN INSTRUMENTS PRODUCTUNKNOWN BUR GUARD| UNKNOWN INSTRUMENTS PRODUCTUNKNOWN BUR