FDA Adverse Event Injury Summary report: N

IDENTITY ADX XL DR

MDR report key: 3925564 · Received January 15, 2014

Report

Report Number
2017865-2014-08612
Event Type
Injury
Date Received
January 15, 2014
Date of Event
March 6, 2012
Manufacturer
ST. JUDE MEDICAL, CRMD
Product Code
LWP
PMA / PMN Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THIS HISTORICAL COMPLAINT IS BEING FILED AS PART OF A RETROSPECTIVE REVIEW OF COMPLAINT FILES IN RESPONSE TO A RECENT FDA INSPECTION. THERE IS NO CHANGE TO THE ACTUAL PERFORMANCE OF THE PRODUCT AND THIS REPORT ONLY REPRESENTS AN ENHANCEMENT TO THE REPORTING CRITERIA GOING FORWARD. EVALUATION DESCRIPTION: FINAL ANALYSIS FOUND THE DEVICE EXHIBITED BACKUP VVI OPERATION DUE TO MULTIPLE FLIPPED BITS IN THE PRODUCT CODE. AFTER THE PRODUCT CODE WAS DOWNLOADED, NORMAL FUNCTION RESUMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYMPTOMATIC PATIENT WITH MUSCLE STIMULATION PRESENTED IN THE OPERATING ROOM. THE PULSE GENERATOR EXHIBITED BACKUP VVI OPERATION. THE PULSE GENERATOR WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
39821 IDENTITY ADX XL DR IMPLANTABLE PACEMAKER PULSE GENERATOR LWP ST. JUDE MEDICAL, CRMD 5386

Patients

Seq Age Sex Outcome Treatment
1 83 YR Required Intervention