PROMUS ELEMENT ?
Report
- Report Number
- 2134265-2014-03937
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- January 10, 2014
- Report Date
- June 12, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE IS A COMBINATION PRODUCT. DEVICE EVALUATED BY MFR.: A VISUAL AND MICROSCOPIC EXAMINATION FOUND THAT THE STENT WAS DAMAGED AT THE DISTAL END. THE STRUTS ON THE MOST DISTAL ROW WERE RAISED AND MISALIGNED. THE INVESTIGATOR NOTED THAT THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. THE TIP WAS SLIGHTLY FLARED. THE HYPOTUBE WAS KINKED AT VARIOUS LOCATIONS ALONG ITS LENGTH. NO OTHER ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/ PROCEDURAL FACTORS. (B)(4).
REPORTABLE BASED ON DEVICE ANALYSIS APPROVED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES WAS ENCOUNTERED. VASCULAR ACCESS WAS OBTAINED VIA FEMORAL ARTERY. THE 100% STENOSED, 20X2.75MM TARGET LESION CONTAINED >=90 DEGREE BEND AND WAS LOCATED IN A SEVERELY TORTUOUS AND SEVERELY CALCIFIED RIGHT CORONARY ARTERY (RCA). A 2.75X20MM PROMUS ELEMENT DRUG-ELUTING STENT WAS SELECTED FOR USE AND ADVANCED BUT FAILED TO CROSS THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND PATIENT'S STATUS WAS STABLE. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED DISTAL STENT DAMAGE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403442 | PROMUS ELEMENT ? | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911320270 | 16003014 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |