FDA Adverse Event Malfunction Summary report: N

MICROFRANCE® INSTRUMENT

MDR report key: 3925354 · Received July 10, 2014

Report

Report Number
9680837-2014-00055
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
June 10, 2014
Report Date
June 10, 2014
Manufacturer
XOMED MICROFRANCE MFG
Product Code
GEI
PMA / PMN Number
K993655
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT: CEV134 - FORCEPS BIPOLAR 360MM MOUIEL, LOT: 09-03, MANUFACTURE DATE: SEPTEMBER 2003. CEV134 - FORCEPS BIPOLAR 360MM MOUIEL, LOT: 121101, MANUFACTURE DATE: NOVEMBER 2012. (B)(4): THE ANALYSIS INDICATES THAT THE BLACK PLASTIC PART IS BURNT BETWEEN THE BIPOLAR CONNECTORS. A LARGE AMOUNT OF SURGERY RESIDUE HAS BEEN FOUND IN THE FORCEPS. THE BURN OF THE PLASTIC PART IS THE CONSEQUENCE OF THE FORMATION OF AN ELECTRIC ARC, PROBABLY DUE TO THE PRESENCE OF HUMIDITY IN THE CONNECTION ZONE (CABLE NOT DRIED / BLOOD / TISSUES). THE CLEANING OF THE FORCEPS IS IN ALL LIKELIHOOD INSUFFICIENT. CEV134 (LOTS: 09-03 AND 121101): THE BLACK PLASTIC PARTS ARE BURNT AT THE CONNECTION. THE INSTRUMENTS HAVE BEEN REPAIRED BY AN EXTERNAL CONTRACTOR (WITH AN UNKNOWN ETCHING). THE BURN OF THE PLASTIC PART IS THE CONSEQUENCE OF THE FORMATION OF AN ELECTRIC ARC, PROBABLY DUE TO THE PRESENCE OF HUMIDITY IN THE CONNECTION ZONE (CABLE NOT DRIED / BLOOD / TISSUES). MOREOVER, THE REPAIR REALIZED BY AN EXTERNAL CONTRACTOR MAY HAVE WEAKENED THE INSTRUMENT AND LED TO THE REPORTED INCIDENT. THIS DEVICE IS USED FOR THERAPEUTIC PURPOSES.

Description of Event or Problem · 1

THREE BIPOLAR FORCEPS (CEV133 ¿ QTY 1; CEV134 ¿ QTY 2) WERE RETURNED FOR REPAIR. THE AVAILABLE INFORMATION INDICATES THAT THE INSTRUMENTS PRESENTED WITH EVIDENCE OF HEAT DAMAGE. THERE WAS NO INJURY REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403667 MICROFRANCE® INSTRUMENT ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES GEI XOMED MICROFRANCE MFG CEV133 120401

Patients

Seq Age Sex Outcome Treatment
1