FDA Adverse Event Malfunction Summary report: N

MUSTANG?

MDR report key: 3925263 · Received July 10, 2014

Report

Report Number
2134265-2014-03957
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
April 7, 2014
Report Date
June 11, 2014
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
FGE
PMA / PMN Number
K110122
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: 18 YEARS OR OLDER. DEVICE EVALUATED BY MFR: A VISUAL AND TACTILE EXAMINATION OF THE RETURNED DEVICE FOUND NO KINKS OR DAMAGE ALONG THE ENTIRE LENGTH OF THE DEVICE. POSITIVE PRESSURE WAS APPLIED AND A LEAK WAS NOTED. A FURTHER MICROSCOPIC ANALYSIS NOTED A LONGITUDINAL TEAR IN THE BALLOON BODY. AN EXAMINATION OF THE BALLOON MATERIAL AND MARKERBANDS IDENTIFIED NO ISSUES WHICH COULD POTENTIALLY HAVE CONTRIBUTED TO THE TEAR. NO OTHER DEFECTS WERE NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/ PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON DEVICE ANALYSIS APPROVED ON (B)(4) 2014. IT WAS REPORTED THAT CROSSING DIFFICULTIES WAS ENCOUNTERED. VASCULAR ACCESS WAS OBTAINED WITH RETROGRADE APPROACH. THE 99% STENOSED TARGET LESION WAS LOCATED IN THE FOREARM SHUNT. FOLLOWING THE INSERTION OF A 5F UNSPECIFIED INTRODUCER SHEATH AND A 35MM UNSPECIFIED GUIDE WIRE, A 5.0 X 40, 40CM MUSTANG BALLOON CATHETER WAS SELECTED FOR USE AND ADVANCED BUT FAILED TO CROSS THE LESION. THE DEVICE WAS REMOVED AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS WAS GOOD. HOWEVER, RETURNED DEVICE ANALYSIS REVEALED BALLOON TORN LONGITUDINALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402955 MUSTANG? CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - GALWAY H74939171050440 0016637850

Patients

Seq Age Sex Outcome Treatment
1 INTRODUCER SHEATH: 5F SHEATH| GUIDEWIRE: 35 WIRE