FDA Adverse Event Malfunction Summary report: N

BATTERY REAMER/DRILL II

MDR report key: 3925215 · Received July 10, 2014

Report

Report Number
3009450871-2014-10242
Event Type
Malfunction
Date Received
July 10, 2014
Report Date
June 12, 2014
Manufacturer
DEPUY SYNTHES POWER TOOLS
Product Code
GEY
PMA / PMN Number
PEXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL COMMON DEVICE NAME: MOQ. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. (B)(6). A REVIEW OF THE DEVICE HISTORY RECORD DID NOT INDICATE ANY CONDITIONS DURING MANUFACTURING OPERATIONS THAT COULD BE RELATED TO THE REPORTED CONDITION. THIS DEVICE PASSED ALL MANUFACTURING AND TEST REQUIREMENTS AT THE TIME OF MANUFACTURE. SERVICE HISTORY REVIEW OF THE DEVICE HAS BEEN PERFORMED. THE REVIEW INDICATES THAT THE DEVICE HAS NOT BEEN SERVICED DURING THE PAST 6 MONTHS. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINT ISSUE IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL NARRATIVE: THE VISUAL INSPECTION OF THE RETURNED DEVICE PERFORMED AS PART OF THE MANUFACTURING INVESTIGATION REPORTED THERE WAS RESIDUAL LIQUID WAS BETWEEN MOTOR AND CONTROL UNIT, PROBABLY FROM THE CLEANING. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

DEVICE REPORT FROM SYNTHES (B)(4) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: IT WAS REPORTED THE DEVICE RUNS WITHOUT PRESSING THE TRIGGER AND CHANGES DIRECTION ON ITS OWN. THIS IS REPORT 1 OF 1 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
405071 BATTERY REAMER/DRILL II MOTOR, SURGICAL INSTRUMENT, AC-POWERED GEY DEPUY SYNTHES POWER TOOLS

Patients

Seq Age Sex Outcome Treatment
1