PANTHERA-PUNCHER 9
Report
- Report Number
- 8043909-2014-00001
- Event Type
- Malfunction
- Date Received
- June 26, 2014
- Date of Event
- May 26, 2014
- Report Date
- June 26, 2014
- Manufacturer
- WALLAC OY
- Product Code
- LXG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AS
- Reporter Occupation
- NOT APPLICABLE
Narratives
OUR INVESTIGATION HAS SHOWN THAT THE IONIZER MAY DECREASE THE REACTION SIGNAL LEVEL OF ASSAYS USING ANTIBODY COATED PLATES (SUCH AS THOSE USED FOR IMMUNOASSAYS). WE HAVE ALSO FOUND THAT UNCOATED PLATES (SUCH AS THOSE USED FOR ENZYMATIC ASSAYS) ARE NOT AFFECTED BY THE PHENOMENON. THE PHENOMENON HAS BEEN DETERMINED TO BE LOCALIZED TO WELLS LOCATED BENEATH OR ADJACENT TO WELLS BENEATH THE IONIZER PINS WHEN THE PANTHERA-PUNCHER 9 IS IDLE OR IN SLEEP MODE. OUR INVESTIGATION INTO THE PHENOMENON HAS DETERMINED CONTRIBUTING FACTORS TO BE THE POSITION OF THE PLATE BENEATH THE IONIZER, THE PERIOD OF TIME THE PLATE IS STATIONARY BENEATH THE IONIZER AND THE VOLTAGE OF THE IONIZER UNIT. THE SOFTWARE CURRENTLY ALLOWS THE IONIZER TO REMAIN ACTIVE WHEN THE PUNCHER IS IDLE OR HAS GONE INTO SLEEP MODE. UNDER THESE CERTAIN CONDITIONS THE IONIZER MAY NEGATIVELY IMPACT SAMPLES RESULTS.
THROUGH A CUSTOMER COMPLAINT REPORTED ON (B)(6) 2014, WE BECAME AWARE THAT EXTENDED EXPOSURE IN CLOSE PROXIMITY TO THE IONIZER PINS OF THE PANTHERA-PUNCHER 9 CAN DECREASE THE REACTION SIGNAL LEVEL OF ASSAYS USING ANTIBODY COATED MICROTITER PLATES (SUCH AS THOSE USED FOR IMMUNOASSAYS). NO FALSE NEGATIVE OR FALSE POSITIVE RESULTS ARE KNOWN TO HAVE BEEN REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 373603 | PANTHERA-PUNCHER 9 | PANTHER PUNCHER | LXG | WALLAC OY | 2081-0010 |