FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3925128 · Received July 10, 2014

Report

Report Number
2531779-2014-19507
Event Type
Malfunction
Date Received
July 10, 2014
Report Date
July 2, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/02/2014 WITH THE FOLLOWING FINDINGS: ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. VISUAL INSPECTION OF THE PUMP FOUND SOME COSMETIC DAMAGES. ON INVESTIGATION, THE PUMP POWERED UP TO A DIM AND DISCOLORED VERIFY SCREEN. THE BATTERY COMPARTMENT WAS CRACKED BETWEEN THE CAP AND CASE SEAL. THE KEYPAD COVER WAS TORN BUT ALL THE BUTTONS RESPONDED PROPERLY. REMOVAL OF THE KEYPAD COVER FOUND CONTAMINATION UNDER ALL THE BUTTON CONTACTS. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION FOUND THAT THE DISPLAY WAS DIM/DISCOLORED, THE BATTERY COMPARTMENT WAS CRACKED, AND THERE WAS CONTAMINATION UNDER THE BUTTON CONTACTS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 07/02/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
404324 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 41 YR