FDA Adverse Event
Malfunction
Summary report: N
ADVANTA 2 BED
MDR report key: 3925093
·
Received June 17, 2014
Report
- Report Number
- 3006697241-2014-00518
- Event Type
- Malfunction
- Date Received
- June 17, 2014
- Date of Event
- May 21, 2014
- Report Date
- May 21, 2014
- Manufacturer
- HILL-ROM DE MEXICO S DE RL DE CV
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND A DEFECTIVE SIDECOM POWER CONTROL BOARD. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2013 AND 2014. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SIDE COM POWER CONTROL BOARD TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED. (B)(4)
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT FROM A HILL-ROM TECHNICIAN STATING THE NURSE CALL FUNCTION IS NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT IN ENGINEERING. THERE WAS NO PT/USER INJURY REPORTED. (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 355776 | ADVANTA 2 BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM DE MEXICO S DE RL DE CV | 1190 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |