FDA Adverse Event
Malfunction
Summary report: N
VERSACARE BED
MDR report key: 3925079
·
Received June 17, 2014
Report
- Report Number
- 1824206-2014-01788
- Event Type
- Malfunction
- Date Received
- June 17, 2014
- Date of Event
- May 21, 2014
- Report Date
- May 21, 2014
- Manufacturer
- HILL-ROM INC
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE HILL-ROM TECHNICIAN FOUND THE BED EXIT ALARM NOT WORKING DUE TO THE SCALE POWER CONTROL BOARD SPEAKER INOPERABLE. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2011-2013. IT IS UNK IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED THE SCALE POWER CONTROL BOARD SPEAKER TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem · 1
HILL-ROM RECEIVED A REPORT THAT A HILL-ROM TECHNICIAN FOUND A NOTE ON THE BED STATING THE BED EXIT ALARM IS NOT WORKING. THE BED WAS LOCATED AT THE ACCOUNT IN STORAGE. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 355944 | VERSACARE BED | AC POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM INC | 3200 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |