AVIATOR ASSY TWO LEVEL PLATE SIZE 34
Report
- Report Number
- 0009617544-2014-00301
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- June 17, 2014
- Report Date
- June 17, 2014
- Manufacturer
- STRYKER SPINE-FRANCE
- Product Code
- KWQ
- PMA / PMN Number
- K083562
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
METHOD: NO DEVICE RETURNED;RESULTS: MANUFACTURING FILES COULD NOT BE REVIEWED BECAUSE NO LOT NUMBER WAS PROVIDED AND NO PART WAS RETURNED FOR VISUAL EXAMINATION. NO SPECIFIC DETAILS WERE PROVIDED CONCERNING THE TECHNIQUE USED THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT (SUCH AS OVER-ARTICULATION OF THE DRILL GUIDE/TAP/SCREW). CONCLUSION: THEREFORE, THE EXACT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE IS MULTIFACTORIAL IN NATURE.
METHOD: DEVICE NOT RETURNED; RESULTS: NO SPECIFIC DETAILS WERE PROVIDED CONCERNING THE TECHNIQUE USED, THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT (SUCH AS OVER-ARTICULATION OF THE DRILL GUIDE/TAP/SCREW). THEREFORE, THE EXACT CAUSE CANNOT BE DETERMINED. THE POTENTIAL CAUSE IS MULTIFACTORIAL IN NATURE. CONCLUSION: THE REPORTED DEVICE WAS CONFIRMED BY THE STRYKER REP TO BE AN AVIATOR ASSY TWO LEVEL PLATE. MANUFACTURING FILES COULD NOT BE REVIEWED BECAUSE NO LOT NUMBER WAS PROVIDED AND NO PART WAS RETURNED FOR VISUAL EXAMINATION. A 10 MINUTE DELAY IN SURGERY RESULTED AND THE PLATE WAS SUCCESSFULLY INSTALLED. THERE WERE NO NEW HAZARD/HARMS IDENTIFIED OR BREACH OF SEVERITY.
IT WAS REPORTED THAT ONE OF THE LOCKING WINGS ON THE AVIATOR CERVICAL PLATE POPPED UP LATERALLY AFTER SCREW INSERTION WITH A SELF - DRILLING 4.0 X 14 SCREW. THE SURGEON WAS ABLE TO USE A PENFIELD FOUR AND A MALLET AND KNOCK IT BACK INTO PLACE AND DID NOT TO REPLACE THE PLATE.
IT WAS REPORTED THAT ONE OF THE LOCKING WINGS ON THE AVIATOR CERVICAL PLATE POPPED UP LATERALLY AFTER SCREW INSERTION WITH A SELF - DRILLING 4.0 X 14 SCREW. THE SURGEON WAS ABLE TO USE A PENFIELD FOUR AND A MALLET AND KNOCK IT BACK INTO PLACE AND DID NOT TO REPLACE THE PLATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403778 | AVIATOR ASSY TWO LEVEL PLATE SIZE 34 | IMPLANT-CERVICAL PLATE | KWQ | STRYKER SPINE-FRANCE | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |