FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3924721 · Received July 10, 2014

Report

Report Number
1416980-2014-21993
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, IT WAS REPORTED THAT THE HOME PATIENT CONNECTED TO THE PATIENT LINE BEFORE THE LINE WAS PROPERLY PRIMED. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

DURING TROUBLESHOOTING FOR A CHECK FINAL LINE ALARM, IT WAS REPORTED THAT THE HOME PATIENT WAS CONNECTED TO THE PATIENT LINE DURING THE PRIMING PHASE OF PERITONEAL DIALYSIS THERAPY. PROPER PROCEDURES WERE REVIEWED. THE CAREGIVER WAS ADVISED TO DISCONTINUE THERAPY BUT THEY INSISTED ON CONTINUING THERAPY WITH THE SAME SUPPLIES. THE CAREGIVER WAS ADVISED TO NOTIFY THEIR NURSE OF THE EVENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403096 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 63 YR HOMECHOICE