FDA Adverse Event Malfunction Summary report: N

VERSACARE BED

MDR report key: 3924700 · Received June 25, 2014

Report

Report Number
1824206-2014-01838
Event Type
Malfunction
Date Received
June 25, 2014
Date of Event
May 28, 2014
Report Date
May 28, 2014
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AK, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE HILL-ROM TECHNICIAN FOUND THE BRAKE/STEER CASTERS NOT HOLDING DUE TO THE INTERNAL LOCKING MECHANISM IS NOT HOLDING PROPERLY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012-2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN REPLACED TWO BRAKE HEAD END CASTERS AND TWO FOOT END STEER CASTERS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKE/STEER CASTERS ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN REHAB ROOM 4. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369808 VERSACARE BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 3200

Patients

Seq Age Sex Outcome Treatment
1