HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-21991
- Event Type
- Malfunction
- Date Received
- July 10, 2014
- Date of Event
- June 15, 2014
- Report Date
- June 15, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, AN INCOMPLETE PRIME WAS REPORTED AND IS A KNOWN CAUSE OF THIS ALARM. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.
IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN SET / LINE) OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING THE INITIAL DRAIN. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. THE HP STATED THAT THERE WERE AIR BUBBLES IN THE PATIENT LINE. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT LINE WAS NOT PROPERLY PRIMED BEFORE CONNECTING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HP TO CYCLE THE POWER ON THE HC TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO START OVER USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 403352 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR | HOMECHOICE |