FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3924679 · Received July 10, 2014

Report

Report Number
1416980-2014-21991
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
June 15, 2014
Report Date
June 15, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, AN INCOMPLETE PRIME WAS REPORTED AND IS A KNOWN CAUSE OF THIS ALARM. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF ADDITIONAL RELEVANT INFORMATION IS OBTAINED, THEN A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR (SE) 2240 ALARM (AIR IN SET / LINE) OCCURRED ON THE HOMECHOICE (HC) DEVICE DURING THE INITIAL DRAIN. THE HOME PATIENT (HP) WAS CONNECTED AT THE TIME OF THE ALARM. THE HP STATED THAT THERE WERE AIR BUBBLES IN THE PATIENT LINE. DURING THE TROUBLESHOOTING, IT WAS DETERMINED THAT THE PATIENT LINE WAS NOT PROPERLY PRIMED BEFORE CONNECTING. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) INSTRUCTED THE HP TO CYCLE THE POWER ON THE HC TO CLEAR THE ALARM. THE TSR ADVISED THE HP TO START OVER USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
403352 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 38 YR HOMECHOICE