ANIMAS VIBE
Report
- Report Number
- 2531779-2014-19481
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- June 19, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
FOLLOW-UP #1: DATE OF SUBMISSION 08/07/2014. DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE TEXT ON THE DISPLAY SCREEN WAS DIM/FADED AND DISCOLORED. UNRELATED TO THE DISPLAY ISSUE, THE BATTERY COMPARTMENT HAD MULTIPLE CRACKS, ONE AT OPENING OF BATTERY CHAMBER EXTENDING DOWN TO THE PRIMARY SEAL AND ONE BELOW THE FINGER PAD EXTENDING DOWN TO THE PRIMARY SEAL. A TEST BATTERY CAP WAS ABLE TO FULLY TIGHTEN UNTIL O-RING WAS SEATED AND CAP WAS FLUSH WITH THE PUMP CASE. THE YELLOW O-RING WAS VISIBLE THROUGH THE CRACK. THERE WAS NO ISSUE WITH A LOSS OF POWER AND NO EVIDENCE OF MOISTURE DAMAGE IN BATTERY COMPARTMENT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401317 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |