FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3924509 · Received July 9, 2014

Report

Report Number
2531779-2014-19477
Event Type
Malfunction
Date Received
July 9, 2014
Report Date
June 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE PUMP POWERED ON TO A DIM DISCOLORED DISPLAY. A BATTERY COMPARTMENT CRACK WAS OBSERVED FROM THE THREAD AREA TO THE CASE SEAL. THE KEYPAD WAS TORN AT THE CONTRAST BUTTON AND WAS PEELING. THE CONTRAST KEY WAS NOT RESPONDING. THE UP AND DOWN KEYS WERE INTERMITTENTLY UNRESPONSIVE. THE CONTRAST BUTTON CONTACT WAS MISSING. CONTAMINATION WAS FOUND UNDERNEATH THE UP AND DOWN KEY CONTACTS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED UNRESPONSIVE KEYPAD BUTTONS WITH CONTAMINATION ON THE KEYPAD CONTACTS AND A DIM AND DISCOLORED DISPLAY SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON (B)(4) 2014.