FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN INFUSION PUMP

MDR report key: 3924504 · Received July 9, 2014

Report

Report Number
2531779-2014-19468
Event Type
Malfunction
Date Received
July 9, 2014
Report Date
June 19, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K042873
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/19/2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS TORN OVER THE DOWN ARROW BUTTON. THE CONTRAST BUTTON WAS INTERMITTENTLY UNRESPONSIVE. CONTAMINATION WAS FOUND UNDERNEATH THE UP AND CONTRAST BUTTON CONTACTS. THE PUMP BOOTED UP TO A DIM AND DISCOLORED DISPLAY SCREEN. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED UNRESPONSIVE KEYPAD BUTTONS WITH CONTAMINATION ON THE KEYPAD CONTACTS IN ADDITION TO DAMAGE TO THE KEYPAD AND A DIM AND DISCOLORED DISPLAY SCREEN. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 06/19/2014.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402083 ANIMAS INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1