FDA Adverse Event Malfunction Summary report: N

5MM MONOPOLAR HANDLE 33CM

MDR report key: 3924498 · Received July 10, 2014

Report

Report Number
0002936485-2014-00461
Event Type
Malfunction
Date Received
July 10, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
STRYKER ENDOSCOPY-SAN JOSE
Product Code
KOG
PMA / PMN Number
K973259
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED. THE MONOPOLAR HANDLE WAS VISUALLY INSPECTED FOR DAMAGES, AND A SCRATCH WAS SEEN NEAR THE DISTAL END. THE HANDLE FAILED INSULSCAN TEST NEAR THE DISTAL END OF THE SHAFT. PROBABLE ROOT CAUSES COULD BE DUE TO USER MISUSE, OR HANDLING DURING CLEANING/REPROCESSING, OR NORMAL WEAR AND TEAR. IN SUM, THE PRODUCT WAS RETURNED AND THE FAILURE MODE WAS CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULATION HAS BEEN COMPROMISED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
405246 5MM MONOPOLAR HANDLE 33CM ENDOSCOPE AND/OR ACCESSORIES KOG STRYKER ENDOSCOPY-SAN JOSE 0647015D

Patients

Seq Age Sex Outcome Treatment