M2A-MAGNUM 42-50 TPR INSRT STD
Report
- Report Number
- 0001825034-2014-06082
- Event Type
- Injury
- Date Received
- July 9, 2014
- Date of Event
- March 20, 2013
- Report Date
- July 8, 2014
- Manufacturer
- BIOMET ORTHOPEDICS
- Product Code
- KWA
- PMA / PMN Number
- PK042037
- Removal / Correction Number
- N/A
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- ATTORNEY
Narratives
CURRENT INFORMATION IS INSUFFICIENT TO PERMIT A CONCLUSION AS TO THE CAUSE OF THE EVENT. REVIEW OF DEVICE HISTORY RECORDS SHOW THAT LOT RELEASED WITH NO RECORDED ANOMALY RELATED TO THE EVENT. THERE ARE WARNINGS IN THE PACKAGE INSERT THAT STATE THAT THIS TYPE OF EVENT CAN OCCUR: UNDER POSSIBLE ADVERSE EFFECTS: "INTRAOPERATIVE OR POSTOPERATIVE BONE FRACTURE AND/OR POSTOPERATIVE PAIN." AND " UNDESIRABLE SHORTENING OF LIMB.¿ THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. THIS REPORT IS NUMBER 4 OF 4 MDRS FILED FOR THE SAME PATIENT (REFERENCE 1825034-2014-02041 /-05112 /-06081 /-06082).
LEGAL COUNSEL FOR PATIENT REPORTED THAT PATIENT UNDERWENT A LEFT TOTAL HIP ARTHROPLASTY ON (B)(6) 2010. SUBSEQUENTLY, THE PATIENT UNDERWENT A REVISION PROCEDURE ON (B)(6) 2013 DUE TO PATIENT ALLEGATIONS OF ELEVATED METAL IONS, LACK OF MOBILITY, PAIN, DYSFUNCTION, SWELLING, INFLAMMATION, LOSS OF RANGE OF MOTION, AND DAMAGE TO SURROUNDING BONE AND TISSUE. THIS REPORT IS BASED ON ALLEGATIONS SET FORTH IN PLAINTIFF¿S COMPLAINT AND THE ALLEGATIONS CONTAINED THEREIN ARE UNVERIFIED. OPERATIVE REPORT RECEIVED NOTED PATIENT UNDERWENT A LEFT HIP REVISION ON (B)(6) 2013 DUE TO PAIN, LOOSENING, STEM SUBSIDENCE AND LEG LENGTH DISCREPANCY. OP REPORT NOTED THE PRESENCE OF SCAR TISSUE, NO EVIDENCE OF METALLOSIS AND THE ACETABULAR CUP WAS STABLE AND REMAINS IMPLANTED. THE MODULAR HEAD, TAPER ADAPTER AND STEM WERE REMOVED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401606 | M2A-MAGNUM 42-50 TPR INSRT STD | PROSTHESIS, HIP | KWA | BIOMET ORTHOPEDICS | N/A | 058120 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Hospitalization| R |