PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-03469
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- June 16, 2014
- Report Date
- June 16, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX
- Reporter Occupation
- PATIENT
Narratives
FAILURE ANALYSIS REVEALED THAT THE INSULIN PUMP ALARMED MOTOR ERROR DURING THE BASIC OCCLUSION TEST AND WAS UNABLE TO PRIME DURING THE PRIME TEST DUE TO A FAULTY FORCE SENSOR. THE MOTOR PASSED MOTOR TEST. UNABLE TO PERFORM THE OCCLUSION TEST DUE TO PRIME ANOMALY. THE DEVICE WAS RECEIVED WITH CRACKED BATTERY TUBE THREADS, RESERVOIR TUBE, RESERVOIR TUBE LIP, MINOR SCRATCHED DISPLAY WINDOW AND MISSING END CAP STICKER.
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
CUSTOMER REPORTED THAT SHE HAD UNEXPLAINED HIGH BLOOD GLUCOSE AND HER INSULIN PUMP DO NOT SEEM TO BE WORKING. TROUBLESHOOTING WAS PERFORMED, AND THE INSULIN PUMP FAILED THE HIGH-PRESSURE TEST TWICE. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401832 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-523LNAS |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR |