FDA Adverse Event
Malfunction
Summary report: N
ISOFLEX S
MDR report key: 3924272
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-15186
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- May 9, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CN
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. ANALYSIS FOUND NORMAL DEVICE CHARACTERISTICS.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING IMPLANT, THE LEAD EXHIBITED UNACCEPTABLE THRESHOLD. THE LEAD WAS NOT IMPLANTED, AND A NEW ONE WAS USED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400625 | ISOFLEX S | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1646T/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |