FDA Adverse Event
Malfunction
Summary report: N
TENDRIL SDX
MDR report key: 3924202
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-15128
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- May 28, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- DTB
- PMA / PMN Number
- P960013
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION AT 39.2CM TO 39.4CM FROM THE CONNECTOR PIN. THE ABRASION WAS CONSISTENT WITH CONSTANT FRICTION TO ANOTHER IMPLANTABLE DEVICE. (B)(6).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 399745 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | DTB | ST. JUDE MEDICAL, INC., CRMD | 1688T/100 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |