FDA Adverse Event Malfunction Summary report: N

TENDRIL SDX

MDR report key: 3924202 · Received July 9, 2014

Report

Report Number
2017865-2014-15128
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
May 28, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DTB
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION AT 39.2CM TO 39.4CM FROM THE CONNECTOR PIN. THE ABRASION WAS CONSISTENT WITH CONSTANT FRICTION TO ANOTHER IMPLANTABLE DEVICE. (B)(6).

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
399745 TENDRIL SDX PERMANENT PACEMAKER ELECTRODE DTB ST. JUDE MEDICAL, INC., CRMD 1688T/100 NA

Patients

Seq Age Sex Outcome Treatment
1