FDA Adverse Event
Malfunction
Summary report: N
TENDRIL SDX
MDR report key: 3924163
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-15083
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- April 14, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION: FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION EXPOSING OUTER COIL AT 48.8 CM TO 49.8 CM FROM CONNECTOR PIN. THE ABRASION IS CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE. (B)(4).
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 400329 | TENDRIL SDX | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1688TC/52 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |