FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3923876 · Received July 9, 2014

Report

Report Number
3004209178-2014-86600
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
June 11, 2014
Report Date
June 11, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP ALARMED DURING THE BASIC OCCLUSION TEST DUE TO PROTRUDED/LOOSE DRIVE SUPPORT DISK. UNABLE TO PERFORM THE OCCLUSION AND EXCESSIVE NO DELIVERY TEST DUE TO PRIME ALARM. THE INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER, CRACKED RESERVOIR TUBE LIP, MINOR SCRATCHED LCD WINDOW, CRACKED BATTERY TUBE THREADS AND SCRATCHED RESERVOIR TUBE WINDOW.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN DOES NOT STOP FLOWING DURING THE PRIME PROCESS. INSULIN SQUIRTED OUT AND THE INSULIN PUMP ALARMED DURING THE PRIME PROCESS. THE BLOOD GLUCOSE READING IS 331 MG/DL. CUSTOMER TREATED WITH MANUAL INJECTION. THE DRIVE SUPPORT IS PROTRUDED. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402067 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-723NAP A3723NAPJ

Patients

Seq Age Sex Outcome Treatment
1 20 YR