FDA Adverse Event Injury Summary report: N

GORE® EXCLUDER® AAA ENDOPROSTHESIS

MDR report key: 3923807 · Received July 9, 2014

Report

Report Number
2017233-2014-00353
Event Type
Injury
Date Received
July 9, 2014
Date of Event
June 13, 2014
Report Date
July 10, 2014
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P020004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

MEDICATIONS: PLAVIX, CRESTOR, LOW-DOSE ASA, LOSARTAN. A REVIEW OF THE MANUFACTURING RECORDS FOR THE DEVICE VERIFIED THE LOT MET ALL PRE-RELEASE SPECIFICATIONS.

Additional Manufacturer Narrative · 1

THE FOLLOWING ADDITIONAL INFORMATION WAS RECEIVED ON 7/10/2014: THE PATIENT WAS TREATED BY EXTENDING THE GRAFT TO THE INFRARENAL LEVEL WITH A 20MM HEMASHIELD. THE EXCLUDER DEVICE WAS LEFT IN PLACE. THE PATIENT RECEIVED ONE UNIT OF BLOOD AND IS CURRENTLY DOING FINE.

Description of Event or Problem · 1

(B)(6) 2013, A PATIENT UNDERWENT TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM WITH GORE® EXCLUDER® AAA ENDOPROSTHESES (RMT311413/10667422). THE PATIENT'S ANATOMY WAS DESCRIBED AS TORTUOUS BUT NO THROMBUS OR CALCIFICATION WAS REPORTED. ON (B)(6) 2014, THE PATIENT PRESENTED EMERGENTLY TO THE HOSPITAL WITH ABDOMINAL PAIN, AND A NON-CONTRAST CT SCAN SHOWED AN IRREGULAR APPEARANCE OF THE RIGHT LIMB OF THE GRAFT, DESCRIBED AS A ¿HALO¿. ADDITIONALLY AN AORTIC NECK IRREGULARITY WAS SEEN AT THE TRANS-RENAL ARTERY LEVEL. LATER THE SAME DAY A CT SCAN WITH CONTRAST WAS PERFORMED. THERE WAS NO VISIBLE LEAK SEEN WITH THIS SCAN AND NO ANEURYSMAL ENLARGEMENT. THE PATIENT WAS TAKEN TO THE PROCEDURAL SUITE WHERE AN EXPLORATORY PROCEDURE WAS UNDERTAKEN. THE PHYSICIAN INDICATED HE PLANNED TO PERFORM A "HYBRID" PROCEDURE DEPENDING ON HIS FINDINGS. IT IS NOT KNOWN WHETHER THIS WAS CARRIED OUT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401957 GORE® EXCLUDER® AAA ENDOPROSTHESIS SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES 10667422

Patients

Seq Age Sex Outcome Treatment
1 64 YR Other| R