FDA Adverse Event
Injury
Summary report: N
NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM.
MDR report key: 3923742
·
Received July 9, 2014
Report
- Report Number
- 6000034-2014-00974
- Event Type
- Injury
- Date Received
- July 9, 2014
- Date of Event
- June 4, 2014
- Report Date
- October 13, 2014
- Manufacturer
- COCHLEAR LTD.
- Product Code
- MCM
- PMA / PMN Number
- 970051
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- AUDIOLOGIST
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Additional Manufacturer Narrative · 1
THIS REPORT IS FILED NOVEMBER 12TH, 2014.
Description of Event or Problem · 1
PER THE CLINIC, THE PATIENT EXPERIENCED NON AUDITORY PAIN/SENSATION WITH AND WITHOUT DEVICE USE AS WELL AS POOR SOUND QUALITY. REPROGRAMMING ATTEMPTS WERE MADE; HOWEVER, THE ISSUE COULD NOT BE RESOLVED. THE DEVICE WAS EXPLANTED (B)(6) 2014, AND THE PATIENT WAS REIMPLANTED WITH A NEW DEVICE DURING THE SAME SURGERY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401800 | NUCLEUS 24 CHANNEL COCHLEAR IMPLANT SYSTEM. | MCM | MCM | COCHLEAR LTD. | CI512 | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 65 YR | Required Intervention |