LIGAMAX CLIP APPLIER
Report
- Report Number
- 3005075853-2014-04785
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- June 11, 2014
- Report Date
- June 23, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K050344
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- OTHER
Narratives
(B)(4). ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE. TO DATE THE DEVICE HAS NOT BEEN RETURNED. IF THE DEVICE OR FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. AT THE TIME OF THIS SUBMISSION, THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS. ATTEMPTS ARE BEING MADE TO OBTAIN THE FOLLOWING INFORMATION. TO DATE NO RESPONSE HAS BEEN PROVIDED. IF FURTHER DETAILS ARE RECEIVED AT A LATER DATE A SUPPLEMENTAL MEDWATCH WILL BE SENT. DID THE CLIP FALL INTO THE PATIENT? IF SO HOW WAS THE CLIP RETRIEVED? WHAT WAS THE SHAPE OF THE CLIP?
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS FOUND THAT THE EL5ML DEVICE WAS RETURNED PARTIALLY CYCLED WITH THE JAWS OPENED. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED TWELVE CONFORMING CLIPS. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE CLIPS DID NOT CLAMP. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401276 | LIGAMAX CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | L4EK5E |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |