FDA Adverse Event
Injury
Summary report: N
TENDRIL ST
MDR report key: 3923351
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-14641
- Event Type
- Injury
- Date Received
- July 9, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- NVN
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. FINAL ANALYSIS FOUND AN INSULATION ABRASION AT 13 CM AND AT 54 CM FROM THE CONNECTOR PIN, WHICH EXPOSED THE OUTER COIL AT BOTH LOCATIONS. THE ABRASION WAS CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE PHYSICIAN SUSPECTED AN INSULATION ANOMALY. THE LEAD WAS EXPLANTED AND REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401251 | TENDRIL ST | PERMANENT PACEMAKER ELECTRODE | NVN | ST. JUDE MEDICAL, INC., CRMD | 1888TC/58 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |