FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 3923351 · Received July 9, 2014

Report

Report Number
2017865-2014-14641
Event Type
Injury
Date Received
July 9, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. EVALUATION DESCRIPTION INCLUDED. FINAL ANALYSIS FOUND AN INSULATION ABRASION AT 13 CM AND AT 54 CM FROM THE CONNECTOR PIN, WHICH EXPOSED THE OUTER COIL AT BOTH LOCATIONS. THE ABRASION WAS CONSISTENT WITH EXPOSURE TO CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PHYSICIAN SUSPECTED AN INSULATION ANOMALY. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401251 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1888TC/58 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention