FDA Adverse Event Injury Summary report: N

TENDRIL ST

MDR report key: 3923281 · Received July 9, 2014

Report

Report Number
2017865-2014-14576
Event Type
Injury
Date Received
July 9, 2014
Date of Event
May 19, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
NVN
PMA / PMN Number
NA
Removal / Correction Number
NA
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FINAL ANALYSIS FOUND AN INSULATION ABRASION EXPOSING THE OUTER COIL AT 18.0 CM TO 19.5 CM FROM THE CONNECTOR PIN. THE ABRASION WAS CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT UPON INTERROGATION THE RIGHT VENTRICULAR LEAD EXHIBITED NOISE. THE LEAD WAS EXPLANTED AND REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
402047 TENDRIL ST PERMANENT PACEMAKER ELECTRODE NVN ST. JUDE MEDICAL, INC., CRMD 1888TC/52 NA

Patients

Seq Age Sex Outcome Treatment
1 69 YR Required Intervention