FDA Adverse Event
Malfunction
Summary report: N
QUICKFLEX XL LV LEAD
MDR report key: 3923200
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-14518
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- April 21, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- OJX
- PMA / PMN Number
- P030054
- Removal / Correction Number
- Z1483
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION EXPOSING THE RING ELECTRODE CABLE LUMEN AT 80.2 CM TO 81.0 CM AND 82.0 CM TO 82.4 CM FROM THE CONNECTOR PIN.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401014 | QUICKFLEX XL LV LEAD | PERMANENT PACEMAKER ELECTRODE | OJX | ST. JUDE MEDICAL, INC., CRMD | 1158T/86 | NA |