FDA Adverse Event Malfunction Summary report: N

QUICKFLEX XL LV LEAD

MDR report key: 3923200 · Received July 9, 2014

Report

Report Number
2017865-2014-14518
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
April 21, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
OJX
PMA / PMN Number
P030054
Removal / Correction Number
Z1483
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND INSULATION ABRASION EXPOSING THE RING ELECTRODE CABLE LUMEN AT 80.2 CM TO 81.0 CM AND 82.0 CM TO 82.4 CM FROM THE CONNECTOR PIN.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
401014 QUICKFLEX XL LV LEAD PERMANENT PACEMAKER ELECTRODE OJX ST. JUDE MEDICAL, INC., CRMD 1158T/86 NA

Patients

Seq Age Sex Outcome Treatment
1