FDA Adverse Event
Malfunction
Summary report: N
QUICK FLEX LV LEAD
MDR report key: 3923193
·
Received July 9, 2014
Report
- Report Number
- 2017865-2014-14510
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Date of Event
- April 10, 2014
- Manufacturer
- ST. JUDE MEDICAL, INC., CRMD
- Product Code
- OJX
- PMA / PMN Number
- P030054
- Removal / Correction Number
- Z1483
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED. NO COMPLAINT WAS RECEIVED WITH RETURN OF DEVICE. FAILURE EVENT OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED AT 65.7 CM TO 67.0 CM AND AT 69.6 CM TO 69.8 CM FROM THE CONNECTOR PIN.
Description of Event or Problem · 1
THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 401938 | QUICK FLEX LV LEAD | PERMANENT PACEMAKER ELECTRODE | OJX | ST. JUDE MEDICAL, INC., CRMD | 1156T/75 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |