FDA Adverse Event Malfunction Summary report: N

ACCENT DR RF

MDR report key: 3923066 · Received July 9, 2014

Report

Report Number
2017865-2014-14405
Event Type
Malfunction
Date Received
July 9, 2014
Date of Event
April 17, 2014
Manufacturer
ST. JUDE MEDICAL, INC., CRMD
Product Code
DXY
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED.

Description of Event or Problem · 1

IT WAS REPORTED THAT AN ASYMPTOMATIC PATIENT PRESENTED FOR ROUTINE FOLLOW UP. UPON INTERROGATION, THE PULSE GENERATOR EXHIBITED NOISE REVERSION; THE DEVICE WAS OVERSENSING ON THE ATRIAL CHANNEL. PROGRAMMING CHANGES WERE NOT MADE. THE DEVICE REMAINED IMPLANTED AND THE PATIENT WOULD BE MONITORED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
399627 ACCENT DR RF IMPLANTABLE PACEMAKER PULSE GENERATOR DXY ST. JUDE MEDICAL, INC., CRMD PM2210 NA

Patients

Seq Age Sex Outcome Treatment
1 86 YR