ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19436
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- July 3, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON 08/13/2014 WITH THE FOLLOWING FINDINGS: THE KEYPAD COVER WAS OBSERVED TO BE TORN NEAR THE DOWN ARROW BUTTON. EVALUATION REVEALED THAT ALL OF THE KEYPAD BUTTONS WERE PROPERLY RESPONSIVE; THE EVENT REPORTED IN THE INITIAL COMPLAINT WAS NOT DUPLICATED. THE PUMP PASSED A LEAK TEST. THE KEYPAD COVER WAS REMOVED, AND EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL OF THE KEY CONTACTS. THE PUMP CASE WAS REMOVED, AND EVIDENCE OF MOISTURE INGRESS WAS NOT OBSERVED ON INTERNAL COMPONENTS. UNRELATED TO THE REPORTED EVENT, THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CNGS PRIOR DMG W/MOIST) ISSUE. IT WAS REPORTED THAT THE DOWN ARROW, UP ARROW, AND OK KEYPAD BUTTONS WERE UNDER RESPONSIVE. THE REPORTER ALLEGED THAT THE KEYPAD WAS TORN/PUNCTURED. IT WAS NOTED THAT THERE WAS MOISTURE/CORROSION IN THE KEYPAD. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 399768 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 14 YR |