ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-19439
- Event Type
- Malfunction
- Date Received
- July 9, 2014
- Report Date
- July 1, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 07/01/2014 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE BATTERY COMPARTMENT WAS CRACKED. UNRELATED TO THIS ISSUE, TESTING REVEALED THE TIME AND DATE RESET TO FACTORY SETTINGS. THE PUMP WAS OPENED AND EVALUATION REVEALED THE INTERNAL CLOCK BATTERY ON THE CIRCUIT BOARD HAD FAILED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED DAMAGE TO THE BATTERY COMPARTMENT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 07/01/2014. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 399427 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 11 YR |